Z-0844-05
Cool/Heat Versatility Main Controller Unit with 7-pin port only (for use with Thermal Resistive Type mattress)
AMF Support Surfaces Inc / initiated 2005-01-27 / Terminated
AMF Support Surfaces Inc began a recall of Cool/Heat Versatility Main Controller Unit with 7-pin port only (for use with Thermal Resistive Type mattress) on . The reason FDA records is "Operator misconnected device power cords causing failure of the device". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0844-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AMF Support Surfaces Inc
- Product
- Cool/Heat Versatility Main Controller Unit with 7-pin port only (for use with Thermal Resistive Type mattress)
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- Operator misconnected device power cords causing failure of the device.
- FDA root cause
- Other
- Action
- Recall letters dated January 28, 2005, wee sent to consignees with instructions that a new control unit was being sent to their attention and directions for return of the previous control units.
- Quantity in commerce
- 6
- Distribution
- IN, MD, ND
- Product codes and lots
- Unit Serial Numbers 1020V0503E00174 1020V0304E00058 1020V0403E00145 1020V0203E00102 1020V0503E00088 1020V0403E00153