Bayer brand Technicon RA/opeRA Systems ISE Fluid Pack for in vitro diagnostic use on the opeRA system only...
Roche Diagnostics Corp. began a recall of Bayer brand Technicon RA/opeRA Systems ISE Fluid Pack for in vitro diagnostic use on the opeRA system only, Manufacturer part number T01-3606-01, Mfd for Bayer... on . The reason FDA records is "A defective seam in the two chamber pouch may result in the mixing of reagents, which would either result in an inability to calibrate the analyzer or would bias patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0608-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Bayer brand Technicon RA/opeRA Systems ISE Fluid Pack for in vitro diagnostic use on the opeRA system only, Manufacturer part number T01-3606-01, Mfd for Bayer Corporation, 511 Benedict Ave, Tarrytown, NY., Roche part number 03361322001.
- Reason for recall
- A defective seam in the two chamber pouch may result in the mixing of reagents, which would either result in an inability to calibrate the analyzer or would bias patient results.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Own name distributor consignees were notified on or about 1/3/05 via letter dated 12/21/04, and were requested to discontinue shipment of the product, notify all of their customers of the recall and to discard product. Roche consignees were notified via letter dated 12/21/04, requested to notify their customers if they had further distributed the product, and instructed to discard all inventory of product.
- Quantity in commerce
- 3181 units
- Distribution
- Nationwide, Argentina, Austria, Brazil, Chile, China, Colombia, Ecuador, France, Germany, Guatemala, India, Indonesia, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Panama, Philippines, Poland, Russia, San Salvador, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom and Uruguay.
- Product codes and lots
- All units.