Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0636-05

Lp(a) Diluent packaged into 60 mL bottles, Catalog #23817, Lot #s 238017-01 & 238017-03; 480 mL bulk catalog...

Diasorin Inc. / initiated 2005-01-10 / Terminated

Diasorin Inc. began a recall of Lp(a) Diluent packaged into 60 mL bottles, Catalog #23817, Lot #s 238017-01 & 238017-03; 480 mL bulk catalog # 95046B, Lot # 238017-02. 60 mL bottles can be... on . The reason FDA records is "QC testing of the Antibody Reagent Set for Lp(a) SPQ Test System it was discovered that the background was failing high of the specification. Through substitution testing it was determined that the Lp(a) diluent, part...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0636-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diasorin Inc.
Product
Lp(a) Diluent packaged into 60 mL bottles, Catalog #23817, Lot #s 238017-01 & 238017-03; 480 mL bulk catalog # 95046B, Lot # 238017-02. 60 mL bottles can be part of DiaSorin Lp(a) Antibody Reagent Kits Catalog # 86084, Lot #s 1144332, 114432A,13656, 113656A, and 113656B for use in DiaSorin Antibody Reagent Set for Lp(a) SPQ Test System .
Product code
DFC Lipoprotein, Low-Density, Antigen, Antiserum, Control
Reason for recall
QC testing of the Antibody Reagent Set for Lp(a) SPQ Test System it was discovered that the background was failing high of the specification. Through substitution testing it was determined that the Lp(a) diluent, part #238017 was the cause of the failing background specification.
FDA root cause
Other
Action
Customers were sent notification letters and instructed to destroy any remaining inventory. Customers were also requested to return a form which indicates number of kits destroyed and a replacement was offered.
Quantity in commerce
631 as individual vials, 1 bulk unit and 401 sold in kits
Distribution
KC, MO, CA, IN, HI, GA, TX, MN, ME, AZ, FL, VA, MI, NJ, NC, NY, CO, WI, MS, BRUSSELS, GREECE, SWEDEN, ENGLAND, SPAIN, CANADA & SWITZERLAND
Product codes and lots
Lp(a) Diluent - 60 mL bottles Catalog #23817, Lot #s 238017-01 & 238017-03; catalog # 95046B, Lot # 238017-02. DiaSorin Lp(a) Antibody Reagent Kits Catalog # 86084, Lot #s 1144332, 114432A,113656, 113656A, and 113656B