Waterlase MD, dental laser unit, Model #6200218, serial numbers 10500210 to 105202680, inclusive

Biolase MG LLC / initiated 2005-01-17 / Terminated

Biolase Technology Inc began a recall of Waterlase MD, dental laser unit, Model #6200218, serial numbers 10500210 to 105202680, inclusive on . The reason FDA records is "Reports that the Waterlase MD, a dental laser unit, water bottle broke under pressure during installation". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0543-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Biolase MG LLC
Product
Waterlase MD, dental laser unit, Model #6200218, serial numbers 10500210 to 105202680, inclusive.
Product code
LYB
Reason for recall
Reports that the Waterlase MD, a dental laser unit, water bottle broke under pressure during installation.
FDA root cause
Other
Action
Water bottles will be replaced by trained Biolase representatives. During that visit, the representative will verify that the bottle pressure is set correctly and the existing bottle is replaced with a new bottle.
Quantity in commerce
206 units
Distribution
Nationwide, Canada, Germany, Ireland, France, United Arab Emirates, Lebanon, Greece, Hong Kong, Argentina, Indonesia, Estonia, Spain, Taiwan
Product codes and lots
Model #6200218; Serial Numbers: 10500210 to 10502680, inclusive.