Z-0543-05
Waterlase MD, dental laser unit, Model #6200218, serial numbers 10500210 to 105202680, inclusive
Biolase MG LLC / initiated 2005-01-17 / Terminated
Biolase Technology Inc began a recall of Waterlase MD, dental laser unit, Model #6200218, serial numbers 10500210 to 105202680, inclusive on . The reason FDA records is "Reports that the Waterlase MD, a dental laser unit, water bottle broke under pressure during installation". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0543-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biolase MG LLC
- Product
- Waterlase MD, dental laser unit, Model #6200218, serial numbers 10500210 to 105202680, inclusive.
- Product code
- LYB
- Reason for recall
- Reports that the Waterlase MD, a dental laser unit, water bottle broke under pressure during installation.
- FDA root cause
- Other
- Action
- Water bottles will be replaced by trained Biolase representatives. During that visit, the representative will verify that the bottle pressure is set correctly and the existing bottle is replaced with a new bottle.
- Quantity in commerce
- 206 units
- Distribution
- Nationwide, Canada, Germany, Ireland, France, United Arab Emirates, Lebanon, Greece, Hong Kong, Argentina, Indonesia, Estonia, Spain, Taiwan
- Product codes and lots
- Model #6200218; Serial Numbers: 10500210 to 10502680, inclusive.