Z-0562-05
Lifepak 500 AED
Physio Control, Inc. / initiated 2005-02-03 / Terminated / root cause: design
Medtronic Emergency Response Systems, Inc. began a recall of LIFEPAK 500 AED on . The reason FDA records is "Potential to not detect patients at the low end of the patient impedance range". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0562-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Physio Control, Inc.
- Product
- Lifepak 500 AED.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Potential to not detect patients at the low end of the patient impedance range.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 2/3/05 the firm issued a 'Medical Device Correction' letter dated January 2005 to their customers. The letter advised customers of the issue and stated a representative from the firm would visit each facility to update affected unit. On 2/25/05 a nationwide press release was issued. On 3/1/05 the firm began telephoning each customer, this was followed by another letter titled MEDICAL DEVICE RECALL. The letter advises of the issue and recommends defibrillators remain in service when no alternative is available until the unit can be corrected. Each unit will be replaced or corrected. On 4/22/05 the firm issued another letter, via certified mail return receipt, expanding their recall. Letter was issued to all customers affected by the expansion of the recall. On 4/22/05 the firm issued a press release advising of the expansion of the recall. The firm's web site also provides a listing of affected units.
- Quantity in commerce
- 2,349
- Distribution
- Devices were distributed to 750 domestic consignees dn 16 international consignees.
- Product codes and lots
- Serial numbers - various, noncontiguous Part numbers - 3005400-XXX, D3005400-XXX, U3005400-XXX
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
