Z-0650-05
Roche ULTIMATE 3 HBA1C test kit for use with MIRA analyzers; catalog number 20755648322
Roche Diagnostics GmbH / initiated 2005-02-15 / Terminated
Roche Diagnostics Corp. began a recall of Roche ULTIMATE 3 HBA1C test kit for use with MIRA analyzers; catalog number 20755648322 on . The reason FDA records is "The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0650-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche ULTIMATE 3 HBA1C test kit for use with MIRA analyzers; catalog number 20755648322.
- Product code
- LCP Assay, Glycosylated Hemoglobin
- Reason for recall
- The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
- FDA root cause
- Other
- Action
- Recall letters dated 2/15/05 were sent to each user instructing customers to run controls before the use of each cassette or kit in a lot, instead of once per lot as previously instructed, and to notify the physician at the facility.
- Quantity in commerce
- 1800 kits.
- Distribution
- Nationwide.
- Product codes and lots
- Lots 646340, 650984, 651640 and 653292.