Z-0650-05

Roche ULTIMATE 3 HBA1C test kit for use with MIRA analyzers; catalog number 20755648322

Roche Diagnostics GmbH / initiated 2005-02-15 / Terminated

Roche Diagnostics Corp. began a recall of Roche ULTIMATE 3 HBA1C test kit for use with MIRA analyzers; catalog number 20755648322 on . The reason FDA records is "The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0650-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche ULTIMATE 3 HBA1C test kit for use with MIRA analyzers; catalog number 20755648322.
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
FDA root cause
Other
Action
Recall letters dated 2/15/05 were sent to each user instructing customers to run controls before the use of each cassette or kit in a lot, instead of once per lot as previously instructed, and to notify the physician at the facility.
Quantity in commerce
1800 kits.
Distribution
Nationwide.
Product codes and lots
Lots 646340, 650984, 651640 and 653292.