LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E
Diasorin Inc. began a recall of LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E on . The reason FDA records is "Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0637-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E.
- Product code
- MRG System, Test, Vitamin D
- Reason for recall
- Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator values.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.
- Quantity in commerce
- 131
- Distribution
- MI, NJ, Germany, Spain and Belgium
- Product codes and lots
- catalog no. 310900, lot 114439E