Z-0637-05

LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E

Diasorin Inc. / initiated 2005-02-14 / Terminated / root cause: manufacturing process

Diasorin Inc. began a recall of LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E on . The reason FDA records is "Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0637-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diasorin Inc.
Product
LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E.
Product code
MRG System, Test, Vitamin D
Reason for recall
Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator values.
Root cause tag
manufacturing process
FDA root cause
Other
Action
DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.
Quantity in commerce
131
Distribution
MI, NJ, Germany, Spain and Belgium
Product codes and lots
catalog no. 310900, lot 114439E