Medtronic Maximo VR, model 7232 Single Chamber Implantable Cardioverter Defibrillator (ICD)devices are tiered...
Medtronic Inc. Cardiac Rhythm Managment began a recall of Medtronic Maximo VR, model 7232 Single Chamber Implantable Cardioverter Defibrillator (ICD)devices are tiered therapy implantable arrhythmia control devices... on . The reason FDA records is "Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0598-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic Maximo VR, model 7232 Single Chamber Implantable Cardioverter Defibrillator (ICD)devices are tiered therapy implantable arrhythmia control devices, which provide automatic detection of ventricular arrhythmias and delivery of user-selected therapies for the detected arrhythmia. Therapies are automatically delivered following detection or can be initiated via programming by the physician. These ICDs are multi-programmable and have the capability of transmitting via telemetry the programmed values, measured and collected data, and event markers.
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs, depletion can take place within a few hours to a few days, after which there is a complete loss of device function.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Medtronic Sales Representatives contacted affected physicians starting 02/10/05 to deliver an Important Patient Management Information. The letter recommended patient managment options including quarterly follow-ups, programming Patient Alert to 'On-High', provide magnet to patients to check device status and inform patients that should they expereicnce warmth in the area surrounding the ICD to seek prompt care.
- Quantity in commerce
- 2,093 total, (1,616 in US & 477 OUS)
- Distribution
- worldwide
- Product codes and lots
- Model 7232 Maximo VR (serial numbers listed include all affected devices but may include some unaffected devices. Patients are recommended to check with their doctor for exact serial numbers affected.) Q-Series: PRN800001A - PRN800024Q. S-series PRN600001S - PRN602430S and PRN604401S - PRN604831S.