Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter...
Baxter Healthcare Corp. began a recall of Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Healthcare Corporation, Deerfield, IL 60015 on . The reason FDA records is "Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0130-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Healthcare Corporation, Deerfield, IL 60015
- Product code
- FRN Pump, Infusion
- Reason for recall
- Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump.
- FDA root cause
- Other
- Action
- Baxter mailed Urgent Device Correction letters dated 2/10/05 to the domestic accounts on 2/10/05 via first class mail, to the attention of the Director of Nursing, advising them of the following hardware problems and software anomalies and providing workarounds for each. There are potential failures of the pump due to upstream occlusion alarms, electro-static discharge, Key-Only security mode (previously covered by letter dated 4/10/02), system error 32 or 33 codes, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump. The accounts were advised that as soon as revised software and hardware is available, Baxter would modify the pumps at no additional charge to the consignee. Any questions were directed to their Baxter Medication Delivery Product Information Center at 1-800-933-0303.
- Quantity in commerce
- 24,009 pumps
- Distribution
- Nationwide including Puerto Rico, and internationally through Baxter subsidiaries in Canada, Chile, Hong Kong, Kuwait and Saudi Arabia.
- Product codes and lots
- product codes 2L3107 and 2L3107R, all serial numbers
FIRST ARTICLE / WEEKLY