Heartsine Samaritan AED Defibrillators - Models: SAM 001, SAM 002, SAM 003. Serial numbers within range of...
Heartsine Technologies Inc began a recall of Heartsine Samaritan AED Defibrillators - Models: SAM 001, SAM 002, SAM 003. Serial numbers within range of 1270 - 2324 (not all are affected) on . The reason FDA records is "The affected Automatic External Defibrillators (AED) annouce a fault warning and shut themselves off before a shock can be delivered to the patient. This is reportedly due to slow capacitor charging rates which are...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0641-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- HeartSine Technologies Ltd
- Product
- Heartsine Samaritan AED Defibrillators - Models: SAM 001, SAM 002, SAM 003. Serial numbers within range of 1270 - 2324 (not all are affected).
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- The affected Automatic External Defibrillators (AED) annouce a fault warning and shut themselves off before a shock can be delivered to the patient. This is reportedly due to slow capacitor charging rates which are interpreted by the AED software as a fault condition.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Recall was initiated by telephone 2/14/2005 with follow-up fax, email or postal mailing using one of several customized letters for distributors or customers. Customers will be sent an upgrade kit by overnite mail and forms to confirm upgrade.
- Quantity in commerce
- 672
- Distribution
- Nationwide to distributors and user facilities. Also to Argentina, Austria, Belgium, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Portugal, Puerto Rico, Singapore, Spain, Switzerland, Thailand, United Kingdom.
- Product codes and lots
- Heartsine Samaritan AED Defibrillators - Models: SAM 001, SAM 002, SAM 003. Serial numbers within range of 1270 - 2324 (not all are affected).