Z-0864-05

Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog...

Zimmer GmbH / initiated 2005-02-09 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part , catalog number 00-4705-003-10 on . The reason FDA records is "The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0864-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.
Distribution
Nationwide and Canada.
Product codes and lots
Chisel lot number 60048084.