Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog...
Zimmer Inc. began a recall of Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part , catalog number 00-4705-003-10 on . The reason FDA records is "The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0864-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- The firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.
- Distribution
- Nationwide and Canada.
- Product codes and lots
- Chisel lot number 60048084.