Z-0662-05

Nellcor brand Pulse Oximeter, Model NPB290, MALLINCKRODT, Made in Ireland, Oxismart, Made in Ireland...

Medtronic, Inc. / initiated 2004-10-18 / Terminated

Nellcor Puritan Bennett began a recall of Nellcor brand Pulse Oximeter, Model NPB290, MALLINCKRODT, Made in Ireland, Oxismart, Made in Ireland; Distributed by Nellcor Puritan Bennett, Inc. Pleasanton... on . The reason FDA records is "The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0662-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Nellcor brand Pulse Oximeter, Model NPB290, MALLINCKRODT, Made in Ireland, Oxismart, Made in Ireland; Distributed by Nellcor Puritan Bennett, Inc. Pleasanton, CA 94588 U.S.A.
Product code
DQA Oximeter
Reason for recall
The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.
FDA root cause
Other
Action
Customers of record were notified of the advisory by letter sent via US Mail on October 18, 2004, via certified mail. A second letter was sent on December 9, 2004, advising customers that upgrades would be available. Upgrades would be available via returns to the firm''s service center, upgrade kits and speaker replacements via self-installation, and a third party service provider.
Quantity in commerce
7,814 units
Distribution
There are 2,950 Direct US and US territory Customers. Product shave also been distributed to Canada, Europe, Japan, Latin America and Australia.
Product codes and lots
All units with serial numbers beginning with G01844386.