Z-0663-05
Nellcor Puritan Bennett brand Pulse Oximeter, Model NPB295, Made in Ireland; Distributed by Nellcor Puritan...
Medtronic, Inc. / initiated 2004-10-18 / Terminated
Nellcor Puritan Bennett began a recall of Nellcor Puritan Bennett brand Pulse Oximeter, Model NPB295, Made in Ireland; Distributed by Nellcor Puritan Bennett, Inc. Pleasanton, CA 94588 U.S.A on . The reason FDA records is "The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0663-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Nellcor Puritan Bennett brand Pulse Oximeter, Model NPB295, Made in Ireland; Distributed by Nellcor Puritan Bennett, Inc. Pleasanton, CA 94588 U.S.A.
- Product code
- DQA Oximeter
- Reason for recall
- The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.
- FDA root cause
- Other
- Action
- Customers of record were notified of the advisory by letter sent via US Mail on October 18, 2004, via certified mail. A second letter was sent on December 9, 2004, advising customers that upgrades would be available. Upgrades would be available via returns to the firm''s service center, upgrade kits and speaker replacements via self-installation, and a third party service provider.
- Quantity in commerce
- 862 units
- Distribution
- There are 2,950 Direct US and US territory Customers. Product shave also been distributed to Canada, Europe, Japan, Latin America and Australia.
- Product codes and lots
- All units with serial numbers beginning with G01844386.
FIRST ARTICLE / WEEKLY