Z-0723-05
Intuiitve Surgical da Vinci 8 mm EndoWrist Curved Scissors, in box. All lots. P/N 400178-01
Intuitive Surgical, Inc. / initiated 2005-03-09 / Terminated
Intuitive Surgical, Inc. began a recall of Intuiitve Surgical da Vinci 8 mm EndoWrist Curved Scissors, in box. All lots. P/N 400178-01 on . The reason FDA records is "Blades on the scissor may break and separate from the main unit as a result of corrosion damage". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0723-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Intuiitve Surgical da Vinci 8 mm EndoWrist Curved Scissors, in box. All lots. P/N 400178-01
- Product code
- LRW Scissors, General, Surgical
- Reason for recall
- Blades on the scissor may break and separate from the main unit as a result of corrosion damage.
- FDA root cause
- Other
- Action
- The firm has sent notification letters to its consignees on 03/09/2005. Consignees are requested to return the feedback form provided by Intuitive Surgical, via Federal Express.
- Quantity in commerce
- 278 units have been distributed.
- Distribution
- The product was distributed to 84 medical facilities, located throughout the US and Switzerland, Belgium, Germany, Australia, Turkey and France.
- Product codes and lots
- FDA Device Listing Number R100529, Model Number 400178-01. 510(k) number K021036.