Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair...
Arizant Inc began a recall of Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair Hugger temperature managemen unit. Not Sterile. Do... on . The reason FDA records is "Air channel seals on a limited number of Bair Hugger Model 555 Pediatric underbody blankets could separate when inflated with air from Bair Hugger warming unit. This separation would create a bulging area in the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0855-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arizant Healthcare, Inc.
- Product
- Bair Hugger Temperature Management Blanket, Model 555 Pediatric Full Access Blanket. Use only with a Bair Hugger temperature managemen unit. Not Sterile. Do not use if primary packaging has been previously opened or is damaged. Arizant Healthcare Inc., 10393 West 70th Street, Eden Prairie, MN 55344 USA.
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- Air channel seals on a limited number of Bair Hugger Model 555 Pediatric underbody blankets could separate when inflated with air from Bair Hugger warming unit. This separation would create a bulging area in the blanket's air channels, which could have a potential to shift the patient's position on the blanket.
- FDA root cause
- Other
- Action
- Recall notification was mailed to customers beginning 03/11/05. Customers were informed of the issue and asked to check their inventory for the affected lots. Affected lots found are to be quarantined and customers are to contact Arizant for instruction on returning affected product and receiving replacement product. Distributors are instructed to examine their inventory for affected lots, discontinue distribution and quarantine the lots. Distributors are asked to contact Arizant for returning affected product. If affected product was distributed, Distributors are to contact Arizant with names, addresses and telephone numbers of their accounts so Arizant can immediately inform those accounts of the recall.
- Quantity in commerce
- 88,320 blankets
- Distribution
- nationwide
- Product codes and lots
- lot numbers ranging from 9631 up to and including 9657
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
