Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF C6121 QTY 1...
Linvatec Corp. began a recall of Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF C6121 QTY 1 3.5mm Bio-Anchor with Disposable... on . The reason FDA records is "Linvatec has determined that there is a possibility that the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1547-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Linvatec Corporation
- Product
- Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF C6121 QTY 1 3.5mm Bio-Anchor with Disposable Driver***STERILE***Linvatec A ConMed Company LINVATEC CORPORATION LARGO, FL 33773 USA***''.
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Linvatec has determined that there is a possibility that the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised sterility of the devices.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Linvatec will issue the recall notifications to all Linvatec direct consignees and customers by mailings using UPS overnight delivery or equivalent method.
- Quantity in commerce
- 1,055 Units
- Distribution
- Product was shipped to consignees and customers (sales representatives, domestic customers, international customers, and international distributors and affiliates).
- Product codes and lots
- Lot # ''s BBB34096, BBB35001, BBB38590, BBB47862, BBB56209, BBB59492, BBB61957, BBB63207, BBB66991, BBB69760, BBB70775.