Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF C6120A QTY...
Linvatec Corp. began a recall of Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF C6120A QTY 1 BIO-ANCHOR, PRE-LOADED 3.5mm x 10.5 mm... on . The reason FDA records is "Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1549-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Linvatec Corporation
- Product
- Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF C6120A QTY 1 BIO-ANCHOR, PRE-LOADED 3.5mm x 10.5 mm ***STERILE ***Linvatec, LINVATEC CORPORATION LARGO, FL 33773 USA.***''
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised sterility of the devices.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Linvatec will issue the recall notifications to all Linvatec direct consignees and customers by mailings using UPS overnight delivery or equivalent method.
- Quantity in commerce
- 1,764 Units
- Distribution
- Product was shipped to consignees and customers (sales representatives, domestic customers, international customers, and international distributors and affiliates).
- Product codes and lots
- Lot #''s BBB31322, BBB33118, BBB35867, BBB36669, BBB37705, BBB39726, BBB40462, BBB45116, BBB46067, BBB46733, BBB52902, BBB55911, BBB57543, BBB59489, BBB61937, BBB63968, BBB66392, BBB67468
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
