Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF*** QTY...
Linvatec Corp. began a recall of Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF*** QTY 1***Hornet Disposable Inserter, BioStinger... on . The reason FDA records is "Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1550-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Linvatec Corporation
- Product
- Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF*** QTY 1***Hornet Disposable Inserter, BioStinger Implant***STERILE***Linvatec, LINVATEC CORPORATION LARGO, FL 33733 USA***''.
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised sterility of the devices.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Linvatec will issue the recall notifications to all Linvatec direct consignees and customers by mailings using UPS overnight delivery or equivalent method.
- Quantity in commerce
- 3,586 Units
- Distribution
- Product was shipped to consignees and customers (sales representatives, domestic customers, international customers, and international distributors and affiliates).
- Product codes and lots
- Lot #''s BBB32927, BBB33173, BBB34093, BBB40588, BBB41400, BBB43500, BBB44122, BBB47021, BBB51947, BBB54961, BBB61003, BBB63978, BBB67469, BBB71590, BBB32928, BBB33170, BBB33172, BBB34090, BBB35000, BBB37713, BBB40589, BBB41401, BBB42354, BBB43059, BBB45125, BBB47861, BBB48837, BBB49882, BBB50769, BBB57155, BBB57550, BBB61002, BBB62845, BBB66407, BBB68489, BBB33174, BBB36672, BBB50768, BBB62844, BBB69766.