Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0656-05

ABSOLOK Extra Absorbable Ligating Clip cartridges, medium/large, code AP300. Ten cartridges are packaged per...

Ethicon Endo-Surgery Inc / initiated 2005-02-28 / Terminated / root cause: packaging

Ethicon Endo-Surgery began a recall of ABSOLOK Extra Absorbable Ligating Clip cartridges, medium/large, code AP300. Ten cartridges are packaged per box on . The reason FDA records is "The product is being recalled because the possibility exists that the sterility of the ligating clips may have been compromised by a tear in the ABSOLOK Extra packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0656-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon Endo-Surgery Inc
Product
ABSOLOK Extra Absorbable Ligating Clip cartridges, medium/large, code AP300. Ten cartridges are packaged per box.
Product code
FZP Clip, Implantable
Reason for recall
The product is being recalled because the possibility exists that the sterility of the ligating clips may have been compromised by a tear in the ABSOLOK Extra packaging.
Root cause tag
packaging
FDA root cause
Other
Action
The recalling firm sent a letter, dated 2/28/05 to consignees.
Quantity in commerce
15,462 boxes.
Distribution
The product was distributed throughout the United States and to foreign customers located in Saudi Arabia, Egypt, United Kingdom, Germany, France, Republic of Panama, Turkey, Hong Kong, Taiwan, Puerto Rico, Spain, Japan, China, Republic of Singapore, Argentina, Italy, New Zealand, Australia, Canada, Czech Republic, Sweden, Portugal, Slovenia, Poland, Hungary, and the United Arab Emirate.
Product codes and lots
Lot: SK2037, exp. date 07/08; SK2055, exp. date 07/08; SL2104, exp. date 07/08; SL2118, exp. date 07/08; SL2445, exp. date 07/08; SM2135, exp. date 07/08; SP2530, exp. date 07/08; SP2711, exp. date 07/08; TA2840, exp. date 01/09; TB2335, exp. date 01/09; TC7957, exp. date 01/09; TD2044, exp. date 01/09; TD2280, exp. date 01/09; TD2664, exp. date 01/09; TD2938, exp. date 01/09; TE7512, exp. date 01/09; TG7713, exp. date 01/09; TG7825, exp. date 01/09; TG7936, exp. date 01/09; TK2477, exp. date 07/09; TK7626, exp. date 07/09; TL2169, exp. date 07/09; TL7265, exp. date 07/09; TP7279, exp. date 07/09; UA2517, exp. date 01/10; UA2747, exp. date 01/10; UA2932, exp. date 01/10; and UA7902, exp. date 01/10.