Z-1046-05
HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003
HeartSine Technologies Ltd / initiated 2005-10-05 / Terminated / root cause: design
Heartsine Technologies Inc began a recall of HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003 on . The reason FDA records is "Immediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts blinking indicating a fault condition". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1046-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- HeartSine Technologies Ltd
- Product
- HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Immediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts blinking indicating a fault condition.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- AED software revision 1.0.20 and higher incorporate software enhancements to make the AED more forgiving for extreme variations in user technique. Firm will provide distributors with field correction notice dated 10/5/05. Customers to be notified by mid-October 2005.
- Quantity in commerce
- 1929 units
- Distribution
- Nationwide, Puerto Rico, Argentina, Australia, Austria, Belgium, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Netherlands, Pakistan, Phillipines, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thailand, UK
- Product codes and lots
- Serial numbers within range of 0000101 and 00003986