Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin...
Del Medical Systems Group began a recall of Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131 on . The reason FDA records is "The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0816-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Del Medical, Inc.
- Product
- Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131
- Product code
- IYB Tube Mount, X-Ray, Diagnostic
- Reason for recall
- The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Del Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
- Quantity in commerce
- 31 stands
- Distribution
- Nationwide, Mexico and Thailand.
- Product codes and lots
- All Phantom Model Tube Stands with squared corners on all four corners of the stand. Catalog #501031, serial numbers 00743-0204, 00778-0204, 00997-0304, 01001-0304, 01103-0304, 01293-0304, 01371-0304, 01425-0404, 01452-0404, 01470-0404, 01475-0404, 01575-0404, 01592-0404, 01700-0404, 01704-0404, 01790-0404, 01841-0404, 01851-0404, 01898-0504, 02019-0504, 02023-0504, 02051-0504, 02096-0504, 02178-0504, 02188-0504, 02310-0604, 02357-0604, 02429-0604, 02517-0604, 02534-0604, 02584-0604, 02671-0604, 02752-0704, 02760-0704, 02776-0704, 02839-0704, 02891-0704, 02928-0704, 02938-0704, 02945-0704, 02959-0704, 02980-0704, 02991-0704, 03076-0704, 03084-0704, 03122-0704, 03375-0804, 03455-0804, 03476-0804, 03518-0804, 03543-0804, 03570-0904, 03579-0904, 03636-0904, 03700-0904, 03708-0904, 03722-0904, 03740-0904, 03952-0904, 04037-0904, 04134-1004, 04183-1004, 04347-1004, 04380-1004, 04407-1004, 04477-1004, 04481-1004, 04491-1004, 04524-1004, 04534-1004, 04592-1104, 04600-1104, 04605-1104, 04648-1104, 04694-1104, 04699-1104, 04703-1104, 0439-1104, 04783-1104, 04862-1104, 05190-1204, 05194-1204, 05198-1204, 05216-1204, 05364-1204, 05414-1204, 05418-1204, 05439-1204, 00283-0105, 00340-0105, 00481-0105, 00571-0205, 00575-0205, 00617-0205, 00649-0205, 00763-0205, 00841-0205, 00858-0205, 00891-0205, 00900-0205, 00913-0205