Z-1544-05

Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, I.V. Systems...

Baxter Healthcare Corporation / initiated 2005-02-25 / Terminated / root cause: labeling and UDI

Baxter Healthcare Corp. began a recall of Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, I.V. Systems Division, product codes 2M8153, 2M8153R, 2M8163, 2M8163R on . The reason FDA records is "Swelling of the sealed lead-acid batteries in the infusion pump can cause internal pump damage, and excessive battery discharge can damage the batteries if the pump is left on battery power for an extended period of...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1544-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, I.V. Systems Division, product codes 2M8153, 2M8153R, 2M8163, 2M8163R
Product code
FRN Pump, Infusion
Reason for recall
Swelling of the sealed lead-acid batteries in the infusion pump can cause internal pump damage, and excessive battery discharge can damage the batteries if the pump is left on battery power for an extended period of time after teh Battery Depleted alarm occurs.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Baxter sent letters dated 02/25/05 to all Colleague infusion pump customers to provide them with important supplemental information to the letter dated 1/21/03 (Z-929-1), regarding sealed lead-acid batteries used in the Colleague family of infusion pumps, dealing with swollen batteries and excessive deep discharge. The 2/25/05 letter reiterated the battery service life, replacement and pump storage conditions which were listed in the 1/21/03 letter, and informed the users that current production pumps have an overcurrent protection circuit in the battery harness to help protect the batteries from overcurrent damage and swelling during charging, and advised them of the availability of the battery harness. The letters also provided information to prevent excessive discharge, and advised them that software updates will become available later in 2005. An Operator's Manual Addendum was also included with the 2/25/05 letter, replacing chapter 7 of the manual dealing with Maintenance and Service. Any questions were directed to Baxter's Medication Delivery Services at 1-800-843-7867.
Quantity in commerce
53,686 pumps
Distribution
Nationwide including Puerto Rico, and internationally to Taiwan, Canada, Chile, Korea, Colombia, New Zealand, China, Hong Kong, Brazil, Turkey and Saudi Arabia.
Product codes and lots
product codes 2M8153 and 2M8153R: all serial numbers below 13120001CT;  product codes 2M8163 and 2M8163R: all serial numbers below 13110338TC