Z-0893-05

AxSYM total B-hCG Reagent is composed of one bottle of Coated Microparticles, one bottle of Alkaline...

Abbott Diagnostic International, Ltd. / initiated 2005-03-24 / Terminated / root cause: manufacturing process

Abbott Health Products, Inc. began a recall of AxSYM total B-hCG Reagent is composed of one bottle of Coated Microparticles, one bottle of Alkaline Phosphatase Conjugate and one bottle of Specimen Diluent on . The reason FDA records is "Abbott Diagnostics is investigation a recent increase in customer complaints for AxSYM calibration error code 1048 and controls out of range when using AxSYM Total B-hCG reagent lots. Testing has shown that protein...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0893-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Diagnostic International, Ltd.
Product
AxSYM total B-hCG Reagent is composed of one bottle of Coated Microparticles, one bottle of Alkaline Phosphatase Conjugate and one bottle of Specimen Diluent.
Product code
JHI Visual, Pregnancy Hcg, Prescription Use
Reason for recall
Abbott Diagnostics is investigation a recent increase in customer complaints for AxSYM calibration error code 1048 and controls out of range when using AxSYM Total B-hCG reagent lots. Testing has shown that protein aggregates are present in the specimen diluents of both IMx and the AxSYM Total B-hCG reagent kits. This is responsible for the calibration errors and shift in control values.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Abbot Health Products, Inc. informed and instructed their customers via Device Recall Letter to Identify if they have used or have inventory of the affected product. Discontinue use of and destroy any remaining inventory of reagent lots listed in the Device Recall letter according to their laboratory procedures. Record the quantity of the recalled reagents currently in inventory in their facility that will be destroyed on the attached reply form and fax the form back to Abbott 1-800-777-0051. (US only) Follow their laboratory''s procedure for communicating this issue to the health care providers they serve.
Quantity in commerce
2496
Distribution
Domestic , US Goverment and International ditribution.
Product codes and lots
2614Q100 (5/13/2005);  26278Q100, 26381Q100 (6/21/2005)