Z-1068-05
Sequoia Oxymatic electronic oxygen conservers. Models OM-302, OM-311 ***and OM-312***
Chad Therapeutics, Inc. / initiated 2005-03-25 / Terminated / root cause: manufacturing process
Chad Therapeutics Inc began a recall of Sequoia Oxymatic electronic oxygen conservers. Models OM-302, OM-311 ***and OM-312*** on . The reason FDA records is ""Shot size" below manufacturing specifications. Risk of less than the specified quantity of oxygen delivered per pulse". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1068-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Chad Therapeutics, Inc.
- Product
- Sequoia Oxymatic electronic oxygen conservers. Models OM-302, OM-311 ***and OM-312***.
- Product code
- BZD Ventilator, Non-Continuous (Respirator)
- Reason for recall
- "Shot size" below manufacturing specifications. Risk of less than the specified quantity of oxygen delivered per pulse.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Firm recalled via email and letter on 3/25/2005. Firm is requesting returns of all units. Recall is on-going and is to the user or patient level.
- Quantity in commerce
- ***244 units***
- Distribution
- Nationwide and to Canada, Sweden, Norway***
- Product codes and lots
- For model #OM-302: Serial #21404202668-21404202680, 21404202682-21404202706, 21404202708-21404202709, 21404202711, 21404202713-21404202715, 21404202722, 21404202725, 21404202728, 21404202730, 21404202733-21404202735, 21404202738-21404202739, 21404202758, 21404202761, 21404202763; For model #OM-311: Serial #21404105027, 21404105030, 2140410503041, 21404105030, 21404105054-21404105055-21404105061, 21404105063-21404105066, 21404105068-21404105070, 21404105073-21404105074, 21404105076, 21404105078-21404105083, 21404105085-21404105086, 21404105089-21404105091, 21404105093-21404105094, 21404105096, 21404105100-21404105102, 21404105105, 21404105107-21404105113, 21404105115-21404105116, 21404105119, 21404105123-21404105124, 21404105127-21404105128, 21404105130-21404105131, 21404105133, 21404105137-21404105139, 21404105142-21404105144, 21404105151-21404105152, 21404105157, 21404105161, 21404105164-21404105168, 21404105170, 21404105172, 21404105178, 21404105180, 21404105182, 21404105185, 21404105194, 21404105201, 21404105204, 21404105206, 21404105209, 21404105211-21404105213, 21404105218, 21404105270-21404105271, 21404105273-21404105274, 21404105276-21404105277, 21404105279; For model #OM-312: Serial #21404301082-21404301159, 21404301161-21404301178