Z-0882-05

Roche brand AccuData GTS System; Catalog number 404, material number 03136833001

Roche Diagnostics GmbH / initiated 2005-04-18 / Terminated / root cause: packaging

Roche Diagnostics Corp. began a recall of Roche brand AccuData GTS System; Catalog number 404, material number 03136833001 on . The reason FDA records is "If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0882-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche brand AccuData GTS System; Catalog number 404, material number 03136833001.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
Root cause tag
packaging
FDA root cause
Other
Action
Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
Distribution
Nationwide and Guam.
Product codes and lots
All units.