Roche brand AccuData GTS/GTS Plus System Replacement Base/Module; Catalog number 3426050, material number...
Roche Diagnostics Corp. began a recall of Roche brand AccuData GTS/GTS Plus System Replacement Base/Module; Catalog number 3426050, material number 03426050001 on . The reason FDA records is "If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0883-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche brand AccuData GTS/GTS Plus System Replacement Base/Module; Catalog number 3426050, material number 03426050001.
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
- Distribution
- Nationwide and Guam.
- Product codes and lots
- All units.