Z-1530-05

Clinac (linear accelerator); Model 21EXS Silhouette Machine (slim version of the Clinac); Distributed by...

Varian Medical Systems, Inc. / initiated 2005-03-30 / Terminated / root cause: design

Varian Medical Systems Inc began a recall of Clinac (linear accelerator); Model 21EXS Silhouette Machine (slim version of the Clinac); Distributed by and/or Manufactured by: Varian Medical Systems, Palo... on . The reason FDA records is "Emergency Off Switch located on the device may not function correctly resulting in the device remaining powered on after the switch is activated". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1530-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Clinac (linear accelerator); Model 21EXS Silhouette Machine (slim version of the Clinac); Distributed by and/or Manufactured by: Varian Medical Systems, Palo Alto, CA 94304.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Emergency Off Switch located on the device may not function correctly resulting in the device remaining powered on after the switch is activated.
Root cause tag
design
FDA root cause
Device Design
Action
On 3/16/05, the firm initiated the recall and their notification was via letters informing its customers of the corrective action.
Quantity in commerce
24 units
Distribution
US distribution is: CA, CT, IL, IN, KY, OH, MI, NC, NJ, NY, SC, TN, TX, VA, & WA, totaling 22 hospitals. Foreign distribution includes: Denmark, Hong Kong, and Japan. It did not involve Defense Supply Center, VA or other Federal Government sales/distribution centers.
Product codes and lots
Affected Serial Numbers: H140001, H140002, H140003, H140004, H140005, H140006, H140007, H140009, H140022, H140010, H14001, H140012, H140013, H140014, H140015, H140016, H140017, H140018, H140019, H140021, H140023, H140024, H140025.