Z-0928-05
BD 25G Spinal Needles. (Spinal Anesthesia Needle)
Becton Dickinson & Co / initiated 2005-03-16 / Terminated
Becton Dickinson & Company began a recall of BD 25G Spinal Needles. (Spinal Anesthesia Needle) on . The reason FDA records is "A significant level of 25G spinal needle hub leakage complaints was observed. Subsequent investigation found the leakage to be attributable to improper formation of the spinal needle hub, specifically an undersized...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0928-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD 25G Spinal Needles. (Spinal Anesthesia Needle)
- Product code
- BSP Needle, Conduction, Anesthetic (W/Wo Introducer)
- Reason for recall
- A significant level of 25G spinal needle hub leakage complaints was observed. Subsequent investigation found the leakage to be attributable to improper formation of the spinal needle hub, specifically an undersized taper and the prescence of a slight sink in the taper surface.
- FDA root cause
- Other
- Action
- The firm sent out notification letters on 3/16/2005.
- Quantity in commerce
- 801580 units
- Distribution
- The firm uses distributors and ships the product directly to over 4000 hospitals and clinics nationwide.
- Product codes and lots
- Reorder Number, Lot Number: 400440, 4275101; 400440, 4282078; 400440, 4317744; 400479, 3196624; 400479, 4161815; 400479, 4254414; 400479, 4324795; 400496, 4173091; 400590, 3191978; 400590, 4121350; 400621, 4146482; 401586, 3182399; 405078, 3196618; 405078, 4169804; 405078, 4194109; 405078, 4268754; 405078, 4282079; 405078, 4324791; 405078, 4349021; 405138, 4173087; 405138, 4181398; 405140, 3224096; 405140, 4110295; 405140, 4237191; 405140, 4295872, 405140, 4314481; 405140, 4336365; 405170, 4201356; 405170, 4201357 405180, 4096782; 405180, 4145523; 405180, 4148675; 405234, 3174061; 405234, 4062807; 405234, 4105111; 405234, 4163586; 405234, 4265265; 405234, 4282715; 405234, 4282716; 405234, 4314480; 405234, 4358850; 409442, 3174062; 409442, 3226365; 409442, 4110275; 409442, 4121348; 409442, 4230810; 409442, 4230813; 409442, 4257418; 409442, 4300308; 409442, 4320423; 409442, 4329215; 409442, 4349941.