FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist...

FHC, Inc. / initiated 2005-03-14 / Terminated / root cause: sterility

FHC, Inc. began a recall of FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN) on . The reason FDA records is "Potential for non-sterile pin to contaminate sterile field". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0895-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FHC, Inc.
Product
FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN)
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Potential for non-sterile pin to contaminate sterile field
Root cause tag
sterility
FDA root cause
Other
Action
FHC notified customers by letter dated 3/14/05 and 3/25/05. The firm has issued a temporary set of instructions which include covering the non-sterile engagement pin with sterile tape and an amended Instruction For Use page to all customers. A redesigned pulley cover which will be sent to all customer when it becomes available in April 2005.
Quantity in commerce
169 units
Distribution
Natiowide
Product codes and lots
All codes