Z-0895-05
FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist...
FHC, Inc. / initiated 2005-03-14 / Terminated / root cause: sterility
FHC, Inc. began a recall of FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN) on . The reason FDA records is "Potential for non-sterile pin to contaminate sterile field". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0895-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FHC, Inc.
- Product
- FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN)
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Potential for non-sterile pin to contaminate sterile field
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- FHC notified customers by letter dated 3/14/05 and 3/25/05. The firm has issued a temporary set of instructions which include covering the non-sterile engagement pin with sterile tape and an amended Instruction For Use page to all customers. A redesigned pulley cover which will be sent to all customer when it becomes available in April 2005.
- Quantity in commerce
- 169 units
- Distribution
- Natiowide
- Product codes and lots
- All codes