Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0913-05

VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label...

Ortho-Clinical Diagnostics, Inc. / initiated 2005-03-15 / Terminated

Ortho-Clinical Diagnostics began a recall of VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label: Ortho-Clinical Diagnostics, Amersham UK on . The reason FDA records is "Potential for positively biased results in QC and patient samples". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0913-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label: Ortho-Clinical Diagnostics, Amersham UK.
Product code
KLI Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine
Reason for recall
Potential for positively biased results in QC and patient samples.
FDA root cause
Other
Action
Letters dated 3/15/2005 sent First Class Mail. Customers are advised product will be replaced. The letter includes modified quality control procedures that will pre-qualify each reagent pack.
Quantity in commerce
Domestic - 1359; International - 840
Distribution
Nationwide distribution. Product shipped worldwide via J&J foreign affiliates.
Product codes and lots
Lot 710, Exp. 7/28/05