Z-0913-05
VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label...
Ortho-Clinical Diagnostics, Inc. / initiated 2005-03-15 / Terminated
Ortho-Clinical Diagnostics began a recall of VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label: Ortho-Clinical Diagnostics, Amersham UK on . The reason FDA records is "Potential for positively biased results in QC and patient samples". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0913-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label: Ortho-Clinical Diagnostics, Amersham UK.
- Product code
- KLI Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine
- Reason for recall
- Potential for positively biased results in QC and patient samples.
- FDA root cause
- Other
- Action
- Letters dated 3/15/2005 sent First Class Mail. Customers are advised product will be replaced. The letter includes modified quality control procedures that will pre-qualify each reagent pack.
- Quantity in commerce
- Domestic - 1359; International - 840
- Distribution
- Nationwide distribution. Product shipped worldwide via J&J foreign affiliates.
- Product codes and lots
- Lot 710, Exp. 7/28/05
FIRST ARTICLE / WEEKLY