Animas Corporation began a recall of Animas IR 1250 Insulin Pump on . The reason FDA records is "pump failed to respond to keypad button pressess after using the food database feature". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0929-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Animas Corp.
- Product
- Animas IR 1250 Insulin Pump.
- Product code
- LZG Pump, Infusion, Insulin
- Reason for recall
- pump failed to respond to keypad button pressess after using the food database feature
- FDA root cause
- Other
- Action
- The recalling firm issued a recall notice dated 4/11/05 to all patients and healthcare providers via first class mail. The letter instructs the patient to reset the pump, thus correcting any problems caused by the software bug, and to discontinue use of the food database. All distributors were contacted by phone on 4/12/05 and instructed to return all pumps in their inventory. On 4/25/05, the recalling began emailing and telephoning patients to arrange for replacement and return of the devices.
- Quantity in commerce
- 2081 units
- Distribution
- The products were shipped to distributors in FL, TX, UT, and VA. The products were also shipped to patients nationwide.
- Product codes and lots
- Model IR1250
FIRST ARTICLE / WEEKLY