Z-0916-05
EBI Ball Tip Guide Wire 3.2mm x 55cm found in the Ankle Arthrodesis Nail Instr., Part Number 469055
Ebi, LLC / initiated 2005-03-16 / Terminated
EBI, L.P. began a recall of EBI Ball Tip Guide Wire 3.2mm x 55cm found in the Ankle Arthrodesis Nail Instr., Part Number 469055 on . The reason FDA records is "The wire appears to be soft and the wire bent during insertion. This causes kinks in which the reamer could not pass over the bent/kinked wire". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0916-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ebi, LLC
- Product
- EBI Ball Tip Guide Wire 3.2mm x 55cm found in the Ankle Arthrodesis Nail Instr., Part Number 469055.
- Product code
- HXI Passer, Wire, Orthopedic
- Reason for recall
- The wire appears to be soft and the wire bent during insertion. This causes kinks in which the reamer could not pass over the bent/kinked wire.
- FDA root cause
- Other
- Action
- The firm distributed a recall letter to their EBI Fracture Sales Representatives and their EBI Trauma Distributors on 3/16/2005.
- Quantity in commerce
- 280 units
- Distribution
- These devices are distributed nationwide to sales representatives who provide the product directly to the hospital/physician at the time of the surgery. The product was also distributed to Italy and Spain. There are no government accounts.
- Product codes and lots
- Lot Number: 069811; 120055; 120056; 120057; 123827; 125637; 125638.