Z-0976-05
ACCU-CHEK Compact Meter; catalog number 3183572
Roche Diagnostics GmbH / initiated 2005-05-18 / Terminated
Roche Diagnostics Corp. began a recall of ACCU-CHEK Compact Meter; catalog number 3183572 on . The reason FDA records is "The meter may self-start without a sample having been applied and report erroneous results without an error message if the nose cover is not properly snapped in place and the undosed test strip is bent upwards while the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0976-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- ACCU-CHEK Compact Meter; catalog number 3183572.
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- The meter may self-start without a sample having been applied and report erroneous results without an error message if the nose cover is not properly snapped in place and the undosed test strip is bent upwards while the meter is waiting for the sample, or if excess control solution or blood has been applied to the test strip.
- FDA root cause
- Other
- Action
- A recall letter dated 5/18/05 was sent to health care professionals informing them of the problem and reqeusting them to provide a letter to their patients alerting them to the issue and how to avoid it. A notice of the problem will be placed in all packages of strips in the future.
- Distribution
- Nationwide.
- Product codes and lots
- All units.