Z-0922-05
Prostate Seeding Needle for Use with the Mick Applicator; an 18 gauge needle, 17 cm long with a white molded...
Argon Medical Devices, Inc / initiated 2005-05-03 / Terminated
Manan Medical Products, Inc. began a recall of Prostate Seeding Needle for Use with the Mick Applicator; an 18 gauge needle, 17 cm long with a white molded plastic hub. The non-sterile prostate seeding... on . The reason FDA records is "The hub of the prostate seeding needle does not release as expected from the applicator as expected during procedures". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0922-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Argon Medical Devices, Inc
- Product
- Prostate Seeding Needle for Use with the Mick Applicator; an 18 gauge needle, 17 cm long with a white molded plastic hub. The non-sterile prostate seeding needles were shipped in bulk for packaging, sterilization and distribution by the customers; the bulk packages were labeled as follows: a) part no. B0140268, Prostate Seeding Set, MICK, InterV - Wheeling, Wheeling, IL 60090 and b) part no. 2380US-18E-17, 18GA x 17CM MK Prostate Seeding Nee, InterV - WHeeling, WHeeling, IL 60090
- Product code
- KXK Source, Brachytherapy, Radionuclide
- Reason for recall
- The hub of the prostate seeding needle does not release as expected from the applicator as expected during procedures.
- FDA root cause
- Other
- Action
- Manan Medical sent recall letters to their bulk needle customers on 5/3/05, informing them of Mick Radio-Nuclear''s recall of their Prostate Seeding Needles manufactured by Manan due to the out of specification needle hub sticking in the applicator. The accounts were informed that the needles manufactured for them were made using the same mold as the Mick needle, and were requested to return the affected lot numbers to Manan for replacement.
- Quantity in commerce
- 22,900 units
- Distribution
- Florida and Georgia
- Product codes and lots
- Part number B0140286, lot numbers 0747, 0839, 0840, 3412, 3413, 3414, 3415, 4078, 4088 and 4089; Part number 2380US-18E-17, lot number 0015