Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0241-06

Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail

Howmedica Osteonics Corp / initiated 2004-11-22 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail on . The reason FDA records is "The Gamma Nail may have been manufactured with an incorrect lag screw hole angulation. Catalog number 3225-0360S, lot code K949396 was manufactured with a lag screw hole angulation of 120 degrees instead of 125 degrees...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0241-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail.
Product code
NDH Nail, Fixation, Bone, Metallic
Reason for recall
The Gamma Nail may have been manufactured with an incorrect lag screw hole angulation. Catalog number 3225-0360S, lot code K949396 was manufactured with a lag screw hole angulation of 120 degrees instead of 125 degrees. Catalog number 3220-0320S, lot code K949408 was manufactured with a lag screw hole angulation of 125 degrees instead of 120 degrees.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Notification letters and Product Accountability forms were sent via FedEx on 11/22/2004 with return receipt.
Quantity in commerce
19
Distribution
The product was sent to Stryker Orthopaedics branches and agents throughout the US. There was one hospital that was a direct account.
Product codes and lots
Catalog 3220-0320S Lot K949408  3225-0360S, lot code K949396