IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by...

Siemens Healthcare Diagnostics Inc. / initiated 2005-04-22 / Terminated

DPC Cirrus began a recall of IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836 on . The reason FDA records is "The Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip of the home sensor". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0994-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip of the home sensor.
FDA root cause
Other
Action
The firm became aware of the problem on 4/19 and conducted tests on 4/22/05. A service alert for field service engineers and Immulite 2000 distributors was prepared on 4/22/05.
Quantity in commerce
There are 941 units in the United States and 3014 units in the international market.
Distribution
worldwide
Product codes and lots
All analyzers; all serial numbers.