Z-0978-05

ACCU-CHEK Instant Meter; Catalog numbers 914 and 924

Roche Diagnostics GmbH / initiated 2005-05-18 / Terminated

Roche Diagnostics Corp. began a recall of ACCU-CHEK Instant Meter; Catalog numbers 914 and 924 on . The reason FDA records is "The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards while the meter is waiting for the sample or if the test strip...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0978-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
ACCU-CHEK Instant Meter; Catalog numbers 914 and 924.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards while the meter is waiting for the sample or if the test strip opening has been previously flooded with excess control solution or blood.
FDA root cause
Other
Action
A recall letter dated 5/18/05 was sent to health care professionals informing them of the problem and reqeusting them to provide a letter to their patients alerting them to the issue and how to avoid it. A notice of the problem will be placed in all packages of strips in the future.
Distribution
Nationwide.
Product codes and lots
All units.