Z-1369-05
Power Processor 3060 Tube Refrigerated Stockyard, P/N 6915556
Siemens Healthcare Diagnostics Inc. / initiated 2005-05-05 / Terminated
Beckman Coulter Inc began a recall of Power Processor 3060 Tube Refrigerated Stockyard, P/N 6915556 on . The reason FDA records is "Device does not maintain a uniform internal temperature within 2-8 C at warmer ambient temperatures". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1369-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Power Processor 3060 Tube Refrigerated Stockyard, P/N 6915556.
- Product code
- LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use
- Reason for recall
- Device does not maintain a uniform internal temperature within 2-8 C at warmer ambient temperatures.
- FDA root cause
- Process change control
- Action
- Recall letters mailed 5/5/2005 request users monitor the temperature variances and evaluate for continued use in thier laboratory.
- Quantity in commerce
- 46
- Distribution
- Nationwide and Canada
- Product codes and lots
- Various.