Mick TP Prostate Seeding Needles, 18 Gauge, Catalog/Model Number MTP-1820-C. --- BRAND NAME: Mick TP...
Mick Radio Nuclear Instr Inc began a recall of Mick TP Prostate Seeding Needles, 18 Gauge, Catalog/Model Number MTP-1820-C. --- BRAND NAME: Mick TP Disposable Needle 18 Gauge. COMMON/USUAL NAME: TP Needle... on . The reason FDA records is "Some of the Mick TP Prostate Seeding Needles do not fit properly into the Mick 200-TPV Applicator. Firm received one complaint during use of the Mick 200 TPV Applicator". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0857-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mick Radio-Nuclear Instruments, Inc.
- Product
- Mick TP Prostate Seeding Needles, 18 Gauge, Catalog/Model Number MTP-1820-C. --- BRAND NAME: Mick TP Disposable Needle 18 Gauge. COMMON/USUAL NAME: TP Needle 18 Gauge. CLASSIFICATION NAME: System, Applicator, Radionuclide, Manual. --- The needle is an accessory to the Mick 200-TPV Applicator and Kit. This one time use needle is used for brachytherapy. The product is sold in cartons with each carton containing a total of 25 needles. Each carton contains 4 packages of 5 needles, and 5 packages containing one needle each.
- Product code
- IWF Needle, Isotope, Gold, Titanium, Platinum
- Reason for recall
- Some of the Mick TP Prostate Seeding Needles do not fit properly into the Mick 200-TPV Applicator. Firm received one complaint during use of the Mick 200 TPV Applicator.
- FDA root cause
- Other
- Action
- Recall Notices (dated 3/30/05) were sent to the consignees by First Class Mail on 3/30/05. Follow-up phone calls to the consignees were initiated on 4/22/05. Recall was expanded for 7 additional lots, and recall notices (dated 8/09/05) were sent to the direct consignees (medical facilities).
- Quantity in commerce
- 327 cartons; RECALL EXPANSION: 926 cartons
- Distribution
- Medical facilities located nationwide, and distributors in Sweden, The Netherlands, Germany and the United Kingdom.
- Product codes and lots
- Lots: M121504-07 (Exp. 12/15/09), M012605-01 (Exp. 1/26/10), M012805-01 (Exp. 1/28/10); --- 8/09/05 RECALL EXPANSION (7 Lots): M110204-01 (Exp. 11/02/09), M110204-04 (Exp. 11/02/09), M110504-02 (Exp. 11/05/09), M120804-01 (Exp. 12/08/09), M120904-01 (Exp. 12/09/09), M121004-03 (Exp. 12/10/09), M121404-01 (Exp. 12/14/09)