Z-0914-05

LeadCare Blood Lead Testing System Kits Catalog Number: 70-2233

Magellan Diagnostics, Inc. / initiated 2005-05-19 / Terminated

Esa Biosciences, Inc. began a recall of LeadCare Blood Lead Testing System Kits Catalog Number: 70-2233 on . The reason FDA records is "Lead test results may be reported lower than the actual value". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0914-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Magellan Diagnostics, Inc.
Product
LeadCare Blood Lead Testing System Kits Catalog Number: 70-2233
Product code
DOF Lead, Atomic Absorption
Reason for recall
Lead test results may be reported lower than the actual value
FDA root cause
Other
Action
ESA is notifying beginning 5/19/05 to Distributors and End Users by letter sent First Class Mail. Additionally Distributors may be notified by phone to ensure receipt of letter. Customers will be asked to return product to ESA for destruction. A fax back form is included with the notification letter. Customers will fax back the number of Test Kits that they require replacement for. ESA will ship replacement kits and include a prepaid return shipping label for the customer to return defective kits to ESA.
Quantity in commerce
14,760 test kits
Distribution
United States Foreign: Mexico, Canada, Indonesia, Palestine, Kingdom of Saudi Arabia, Columbia, Switzerland, Romania, West Indies, Italy, Japan, Turkey, United Kingdom, Phillipines, Singapore, Jamaica, Uraguay, South Korea, Peru, Tanzia, South Africa, Quatar, Croatia, Zambia, China, Germany
Product codes and lots
Lot CA6 Expiration date 2/5/05 Lot CA7 Expiration date 2/06/05 Lot CA8 Expiration date 5/18/05 Lot CA9 Expiration date 10/5/05 Lot CCA Expiration date 12/11/05 Lot CCE Expiration date 6/17/06 Lot CCF Expiration date 7/11/06 Lot CCH Expiration date 7/12/06