Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1370-05

ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter

Roche Diagnostics GmbH / initiated 2005-06-07 / Terminated

Roche Diagnostics Corp. began a recall of ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter on . The reason FDA records is "The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1370-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
Product code
NBW System, Test, Blood Glucose, Over The Counter
Reason for recall
The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample.
FDA root cause
Other
Action
Recall letters dated 6/7/05 were sent to all customers. Distributors were requested to send a copy of the notification to all customers of test strips used with the product. Users are instructed to never operate the meter without the optics cover in place, and this information will be included in test strip boxes.
Quantity in commerce
2490
Distribution
Nationwide.
Product codes and lots
All units.