Z-1370-05
ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter
Roche Diagnostics GmbH / initiated 2005-06-07 / Terminated
Roche Diagnostics Corp. began a recall of ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter on . The reason FDA records is "The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1370-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample.
- FDA root cause
- Other
- Action
- Recall letters dated 6/7/05 were sent to all customers. Distributors were requested to send a copy of the notification to all customers of test strips used with the product. Users are instructed to never operate the meter without the optics cover in place, and this information will be included in test strip boxes.
- Quantity in commerce
- 2490
- Distribution
- Nationwide.
- Product codes and lots
- All units.