Hill-Rom, Inc. began a recall of Hill-Rom Century+ Bed, Model P1400 on . The reason FDA records is "Sleep bed pivot bolts may break, allowing the head section to shift or to drop to one side". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0948-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Hill-Rom Century+ Bed, Model P1400.
- Product code
- FNL Bed, Ac-Powered Adjustable Hospital
- Reason for recall
- Sleep bed pivot bolts may break, allowing the head section to shift or to drop to one side.
- FDA root cause
- Other
- Action
- Consignees were notified via letter dated 5/20/05 that the firm will visit consignees to make corrections to the beds.
- Quantity in commerce
- 218
- Distribution
- Nationwide (California, District of Columbia, Florida, Massachusetts, Montana, Nevada, New Jersey, New York, Pennsylvania, Texas, Virginia), Brazil and Canada.
- Product codes and lots
- Beds built between March 2005 and May 2005.