Z-0915-05
ACL Futura, ACL Advance and ACL TOP Coagulation Analyzers used with the HemosIL RecombiPlasTin reagent
Instrumentation Laboratory Co. / initiated 2005-05-19 / Terminated
Instrumentation Laboratory Co. began a recall of ACL Futura, ACL Advance and ACL TOP Coagulation Analyzers used with the HemosIL RecombiPlasTin reagent on . The reason FDA records is "Patient prothrombin time (PT) may report low for individuals on anticoaglant therapy". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0915-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Instrumentation Laboratory Co.
- Product
- ACL Futura, ACL Advance and ACL TOP Coagulation Analyzers used with the HemosIL RecombiPlasTin reagent
- Product code
- JPA System, Multipurpose For In Vitro Coagulation Studies
- Reason for recall
- Patient prothrombin time (PT) may report low for individuals on anticoaglant therapy
- FDA root cause
- Other
- Action
- Instrumentation Laboratory issued an 'Urgent Product Notification' on 5/19/2005 to Beckman Coulter, Miami FL (DIstributor) to provide the written notification and advise customers to review PT curves an any patient on anticagulant treatment who reports a low result , and manually view the curve where a PT is flagged. A software upgrade will be availabel in July 2005.
- Quantity in commerce
- 617 units
- Distribution
- Nationwide Canada
- Product codes and lots
- Models: 290001 999001 with SN''s beginning with 05041164 280000 with S/N''s beginning with 0505354