Z-0937-05
Potassium Electrode Tip; Part No.s: 668281 Tip Assembly- Synchron Family 669117 Packaged- Synchron Family...
Siemens Healthcare Diagnostics Inc. / initiated 2005-05-20 / Terminated / root cause: manufacturing process
Beckman Coulter Inc began a recall of Potassium Electrode Tip; Part No.s: 668281 Tip Assembly- Synchron Family 669117 Packaged- Synchron Family 441104 Electrode Assembly- EL-ISE 445604 Electrode... on . The reason FDA records is "Does not meet performance criteria and exhibits upward trend of result recoveries during calibration period". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0937-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Potassium Electrode Tip; Part No.s: 668281 Tip Assembly- Synchron Family 669117 Packaged- Synchron Family 441104 Electrode Assembly- EL-ISE 445604 Electrode Packaged- EL-ISE
- Product code
- CEM Electrode, Ion Specific, Potassium
- Reason for recall
- Does not meet performance criteria and exhibits upward trend of result recoveries during calibration period.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Firm sent letters by US Mail requesting that users discard thier probes.
- Quantity in commerce
- 2686
- Distribution
- United States and Canada
- Product codes and lots
- Multiple lots are involved. |SYNCRHON: | | | |Lot 446 ¿ Manufactured 20 NOV 2004* |231 |N/A** | |Lot 450 ¿ Manufactured 12 DEC 2004 |880 | | |Lot 501 ¿ Manufactured 03 JAN 2005 |30 | | |Lot 504 ¿ Manufactured 24 JAN 2005 |698 | | |Lot 507 ¿ Manufactured 21 FEB 2005 |400 | | |Lot 508 ¿ Manufactured 28 FEB 2005 |130 | | |Lot 512 ¿ Manufactured 28 MAR 2005 |1643 | | |Lot 517 ¿ Manufactured 02 MAY 2005 |2020 | | | | | | |EL-ISE: | | | |Lot 0446 ¿ Manufactured 20 NOV 2004* |20 |N/A** | |Lot 0504 ¿ Manufactured 24 JAN 2005 |30 | | |Lot 0508 ¿ Manufactured 28 FEB 2005 |30 | | |---------------------------------------------+---------------+---------------------|