Z-0944-05
Vaxcel Plastic Low Profile Port (non-valved) chest port, with 6Fr Polyurethane Catheter, Catalog 45-333...
Boston Scientific Corporation / initiated 2005-06-08 / Terminated
Boston Scientific Corporation began a recall of Vaxcel Plastic Low Profile Port (non-valved) chest port, with 6Fr Polyurethane Catheter, Catalog 45-333, Material #M001453330. Firm on label: Boston Scientific... on . The reason FDA records is "Potential for port housing separation after implantation". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0944-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Vaxcel Plastic Low Profile Port (non-valved) chest port, with 6Fr Polyurethane Catheter, Catalog 45-333, Material #M001453330. Firm on label: Boston Scientific *** Manufactured at: 10 Glens Falls Technical Park, Glens Falls, NY 12801.
- Product code
- LLD Port & Catheter, Implanted, Subcutaneous, Intraperitoneal
- Reason for recall
- Potential for port housing separation after implantation.
- FDA root cause
- Other
- Action
- A recall package issued to direct accounts on 6/8/05. Hospitals are instructed to return product in inventory, notify physicians, evaluate patients and explant devices whenever possible.
- Distribution
- Product shipped to 56 medical facilities throughout the U.S.
- Product codes and lots
- 914663, 914663D, 924196, 927904, 973169, 994958, 1000904, 1000905, 1003264, 1017119, 1018949, 1023552, 1033976, 1037228, 1037879.