Z-1053-05
BD Syringes - #309585 -- Piston Syringe, Syringe 3ML LL Syringe
Becton Dickinson & Co / initiated 2005-05-25 / Terminated / root cause: sterility
Becton Dickinson & Company began a recall of BD Syringes - #309585 -- Piston Syringe, Syringe 3ML LL Syringe on . The reason FDA records is "The firm believes that some of the syringes in the affected lots may not have received the intended sterilization dose". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1053-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Syringes - #309585 -- Piston Syringe, Syringe 3ML LL Syringe.
- Product code
- FMF Syringe, Piston
- Reason for recall
- The firm believes that some of the syringes in the affected lots may not have received the intended sterilization dose.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- BD contacted their customers via phone and overnight mail letter . End users were notified by overnight mail with confirmation of delivery.
- Quantity in commerce
- 691200 units
- Distribution
- BD distributed the syringes to 18 wholesalers/direct accounts nationwide. These further distributed the products to hospitals and clinics.
- Product codes and lots
- Affected Lot 309585 Syringe 3ML LL Syringe only Lot 5064003