Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0977-05

ACCU-CHEK Compact Care Kit; catalog number 3149137

Roche Diagnostics GmbH / initiated 2005-05-18 / Terminated

Roche Diagnostics Corp. began a recall of ACCU-CHEK Compact Care Kit; catalog number 3149137 on . The reason FDA records is "The meter may self-start without a sample having been applied and report erroneous results without an error message if the nose cover is not properly snapped in place and the undosed test strip is bent upwards while the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0977-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
ACCU-CHEK Compact Care Kit; catalog number 3149137.
Product code
NBW System, Test, Blood Glucose, Over The Counter
Reason for recall
The meter may self-start without a sample having been applied and report erroneous results without an error message if the nose cover is not properly snapped in place and the undosed test strip is bent upwards while the meter is waiting for the sample, or if excess control solution or blood has been applied to the test strip.
FDA root cause
Other
Action
A recall letter dated 5/18/05 was sent to health care professionals informing them of the problem and reqeusting them to provide a letter to their patients alerting them to the issue and how to avoid it. A notice of the problem will be placed in all packages of strips in the future.
Distribution
Nationwide.
Product codes and lots
All units.