Z-1024-05
Depuy brand LCS Complete metal backed patella, porocoat, STD; REF 1294-09-640
Depuy Orthopaedics Inc. / initiated 2005-06-07 / Terminated / root cause: manufacturing process
Depuy Orthopaedics, Inc. began a recall of Depuy brand LCS Complete metal backed patella, porocoat, STD; REF 1294-09-640 on . The reason FDA records is "There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1024-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- Depuy brand LCS Complete metal backed patella, porocoat, STD; REF 1294-09-640.
- Product code
- NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
- Reason for recall
- There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Hospitals were notified via letter dated June 10, 2005 requesting that they quarantine product on hand and send in a response card to the firm.
- Distribution
- Nationwide and international.
- Product codes and lots
- Lots 1831720, 1831721, 1837366, 1837367, 1840603, 1845651 and 1855830.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.