Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1026-05

Depuy brand LCS Complete metal backed patella, porocoat, LRG; REF 1294-09-660

Depuy Orthopaedics Inc. / initiated 2005-06-07 / Terminated / root cause: manufacturing process

Depuy Orthopaedics, Inc. began a recall of Depuy brand LCS Complete metal backed patella, porocoat, LRG; REF 1294-09-660 on . The reason FDA records is "There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1026-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Depuy brand LCS Complete metal backed patella, porocoat, LRG; REF 1294-09-660.
Product code
NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
Reason for recall
There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Hospitals were notified via letter dated June 10, 2005 requesting that they quarantine product on hand and send in a response card to the firm.
Distribution
Nationwide and international.
Product codes and lots
Lots 1829092, 1831726, 1837370, 1845654, 1848385 and 1855839.