Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1120-05

Depuy LCS Complete RP insert; polyethylene tibial insert MED, 25 mm; Ref 1294-05-325

Depuy Orthopaedics Inc. / initiated 2005-05-26 / Terminated / root cause: packaging

Depuy Orthopaedics, Inc. began a recall of Depuy LCS Complete RP insert; polyethylene tibial insert MED, 25 mm; Ref 1294-05-325 on . The reason FDA records is "An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1120-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Depuy LCS Complete RP insert; polyethylene tibial insert MED, 25 mm; Ref 1294-05-325.
Product code
NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
Reason for recall
An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root cause tag
packaging
FDA root cause
Other
Quantity in commerce
15
Distribution
Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
Product codes and lots
Lots X26BW4 and YJ5BJ4.